Starter Kit Instruction Manual Maintenance 47
Sensing level and range adjustement
[4] Replace the cable connected to J8 (4) with
the 100Ω 1% reference resistor in the follow-
ing way:
Connect one side of the resistors to pins 1 a nd
2, the other side to pins 3 and 4.
[5] Measure the voltage on Zn3/R10 (3).
Note: The value should be:
- V = 1.22 to 1.25V.
[6] Meas ure the voltage at the intersection of
R19/R35 (5).
Note: The value should be: V=100mV (ad-
justable with the potent iometer P1 ( 2)).
Note: If the range is too small, replace the
resistor R1 (1) = 1.3k with 1.4k.
[7] Set the se lec tor (9) to display Real
C.
[8] Adjust the trimmer P6 (6) in order to ob-
tain the value of 000
C on the screen (8).
[9] Vary the Temperature Reference by using
the Set Temperature 5 turns potentiometer
(7) located on the front panel.
Note: The range displayed should be -074 to
+ 074.
1
2
3
4
5
6
Fig.39: TCU 151 main board
TCU151
T
e
m
p
e
r
a
t
u
r
e
C
o
n
t
r
o
l
l
e
r
A
l
a
r
m
R
e
s
e
t
E
x
t
e
r
n
a
l
I
n
t
e
r
n
a
l
R
e
a
l
I
R
e
a
l
C
S
e
t
t
i
n
g
C
S
e
t
t
i
n
g
+
I
S
e
t
t
i
n
g
I
+
7
8
9
Fig.40: TCU151 front panel
Zobrazit stránku 47
1 2 ... 43 44 45 46 47 48 49 50 51 52 53 ... 59 60

Komentáře k této Příručce

Žádné komentáře